Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

ET1; PPET1; Preproendothelin-1; Big endothelin-1

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 86-93 89
EDTA plasma(n=5) 87-101 92
heparin plasma(n=5) 89-104 97

Precision of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 92-105% 98-105% 93-105% 96-103%
EDTA plasma(n=5) 98-105% 80-103% 89-97% 83-90%
heparin plasma(n=5) 95-102% 99-105% 80-101% 85-101%

Stability of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards,Labeled antigen and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest.A competitive inhibition reaction is launched between biotin labeled analytes of interest and unlabeled analytes of interest (Standards or samples) with the pre-coated antibody specific to analytes of interest. Following a wash to remove any unbound substances, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE) is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. The MFI developed is reverse proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
RPA482Ra01 Recombinant Endothelin 1 (EDN1) Positive Control; Immunogen; SDS-PAGE; WB.
APA482Ra01 Active Endothelin 1 (EDN1) Cell culture; Activity Assays.
CPA482Ra11 BSA Conjugated Endothelin 1 (EDN1) Immunogen; SDS-PAGE; WB.
CPA482Ra21 OVA Conjugated Endothelin 1 (EDN1) Immunogen; SDS-PAGE; WB.
PAA482Ra08 Polyclonal Antibody to Endothelin 1 (EDN1) WB; IHC; ICC; IP.
PAA482Ra01 Polyclonal Antibody to Endothelin 1 (EDN1) WB; IHC; ICC; IP.
LAA482Ra81 FITC-Linked Polyclonal Antibody to Endothelin 1 (EDN1) WB; IHC; ICC; IF.
LAA482Ra71 Biotin-Linked Polyclonal Antibody to Endothelin 1 (EDN1) WB; IHC; ICC.
MAA482Ra21 Monoclonal Antibody to Endothelin 1 (EDN1) WB; IHC; ICC; IP.
CEA482Ra ELISA Kit for Endothelin 1 (EDN1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA482Ra Multiplex Assay Kit for Endothelin 1 (EDN1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA482Ra11 ELISA Kit DIY Materials for Endothelin 1 (EDN1) Main materials for "Do It (ELISA Kit) Yourself".