ELISA Kit for Motilin (MTL)

MLN; MAP; Promotilin; Motilin-associated peptide

  • ELISA Kit for Motilin (MTL) Packages (Simulation)
  • ELISA Kit for Motilin (MTL) Packages (Simulation)
  • ELISA Kit for Motilin (MTL) Results demonstration
  • CEA575Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Motilin (MTL)

This assay has high sensitivity and excellent specificity for detection of Motilin (MTL).
No significant cross-reactivity or interference between Motilin (MTL) and analogues was observed.

Recovery of the ELISA Kit for Motilin (MTL)

Matrices listed below were spiked with certain level of recombinant Motilin (MTL) and the recovery rates were calculated by comparing the measured value to the expected amount of Motilin (MTL) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 82-95 89
EDTA plasma(n=5) 87-94 90
heparin plasma(n=5) 83-95 88

Precision of the ELISA Kit for Motilin (MTL)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Motilin (MTL) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Motilin (MTL) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Motilin (MTL)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Motilin (MTL) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-101% 85-92% 85-95% 82-102%
EDTA plasma(n=5) 99-105% 81-93% 99-105% 98-105%
heparin plasma(n=5) 97-104% 81-96% 96-105% 80-101%

Stability of the ELISA Kit for Motilin (MTL)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Motilin (MTL)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Test principle of the ELISA Kit for Motilin (MTL)

This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Motilin (MTL) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Motilin (MTL) and unlabeled Motilin (MTL) (Standards or samples) with the pre-coated antibody specific to Motilin (MTL). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Motilin (MTL) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Motilin (MTL) in the sample.

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPA575Hu02 Recombinant Motilin (MTL) Positive Control; Immunogen; SDS-PAGE; WB.
RPA575Hu01 Recombinant Motilin (MTL) Positive Control; Immunogen; SDS-PAGE; WB.
CPA575Hu21 OVA Conjugated Motilin (MTL) Immunogen; SDS-PAGE; WB.
CPA575Hu32 KLH Conjugated Motilin (MTL) Immunogen; SDS-PAGE; WB.
PAA575Hu02 Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
PAA575Hu08 Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
PAA575Hu09 Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
PAA575Hu01 Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
LAA575Hu71 Biotin-Linked Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC.
MAA575Hu21 Monoclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
MAA575Hu22 Monoclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
LAA575Hu72 Biotin-Linked Monoclonal Antibody to Motilin (MTL) WB; IHC; ICC.
CEA575Hu ELISA Kit for Motilin (MTL) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA575Hu Multiplex Assay Kit for Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA575Hu11 ELISA Kit DIY Materials for Motilin (MTL) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK