ELISA Kit for Acetylcholinesterase (ACHE)

ARAChE; N-AChE; YT

  • ELISA Kit for Acetylcholinesterase (ACHE) Packages (Simulation)
  • ELISA Kit for Acetylcholinesterase (ACHE) Packages (Simulation)
  • ELISA Kit for Acetylcholinesterase (ACHE) Results demonstration
  • SEB447Mu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Acetylcholinesterase (ACHE)

This assay has high sensitivity and excellent specificity for detection of Acetylcholinesterase (ACHE).
No significant cross-reactivity or interference between Acetylcholinesterase (ACHE) and analogues was observed.

Precision of the ELISA Kit for Acetylcholinesterase (ACHE)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Acetylcholinesterase (ACHE) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Acetylcholinesterase (ACHE) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability of the ELISA Kit for Acetylcholinesterase (ACHE)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Acetylcholinesterase (ACHE)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the ELISA Kit for Acetylcholinesterase (ACHE)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Acetylcholinesterase (ACHE). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Acetylcholinesterase (ACHE). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Acetylcholinesterase (ACHE), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Acetylcholinesterase (ACHE) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Mus musculus (Mouse) Applications (RESEARCH USE ONLY!)
RPB447Mu01 Recombinant Acetylcholinesterase (ACHE) Positive Control; Immunogen; SDS-PAGE; WB.
RPB447Mu02 Recombinant Acetylcholinesterase (ACHE) Positive Control; Immunogen; SDS-PAGE; WB.
EPB447Mu61 Eukaryotic Acetylcholinesterase (ACHE) Positive Control; Immunogen; SDS-PAGE; WB.
PAB447Mu02 Polyclonal Antibody to Acetylcholinesterase (ACHE) WB; IHC; ICC; IP.
PAB447Mu01 Polyclonal Antibody to Acetylcholinesterase (ACHE) WB; IHC; ICC; IP.
LAB447Mu71 Biotin-Linked Polyclonal Antibody to Acetylcholinesterase (ACHE) WB; IHC; ICC.
SEB447Mu ELISA Kit for Acetylcholinesterase (ACHE) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB447Mu Multiplex Assay Kit for Acetylcholinesterase (ACHE) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
GUESTBOOK