CLIA Kit for Aldosterone (ALD)

  • CLIA Kit for Aldosterone (ALD) Packages (Simulation)
  • CLIA Kit for Aldosterone (ALD) Packages (Simulation)
  • CLIA Kit for Aldosterone (ALD) Results demonstration
  • CCA911Ge.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Aldosterone (ALD)

This assay has high sensitivity and excellent specificity for detection of Aldosterone (ALD).
No significant cross-reactivity or interference between Aldosterone (ALD) and analogues was observed.

Recovery of the CLIA Kit for Aldosterone (ALD)

Matrices listed below were spiked with certain level of Aldosterone (ALD) and the recovery rates were calculated by comparing the measured value to the expected amount of Aldosterone (ALD) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-102 99
EDTA plasma(n=5) 92-99 95
heparin plasma(n=5) 91-103 95

Precision of the CLIA Kit for Aldosterone (ALD)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Aldosterone (ALD) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Aldosterone (ALD) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Aldosterone (ALD)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Aldosterone (ALD) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-101% 87-96% 97-105% 86-93%
EDTA plasma(n=5) 88-99% 92-105% 86-94% 78-101%
heparin plasma(n=5) 90-101% 81-105% 96-105% 81-95%

Stability of the CLIA Kit for Aldosterone (ALD)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Aldosterone (ALD)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
7. Read RLU value immediately.

Test principle of the CLIA Kit for Aldosterone (ALD)

The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Aldosterone (ALD). A competitive inhibition reaction is launched between biotin labeled Aldosterone (ALD) and unlabeled Aldosterone (ALD) (Standards or samples) with the pre-coated antibody specific to Aldosterone (ALD). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Aldosterone (ALD) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Aldosterone (ALD) level in the sample or standard.

Related products

Catalog No. Organism species: Pan-species (General) Applications (RESEARCH USE ONLY!)
MAA911Ge21 Monoclonal Antibody to Aldosterone (ALD) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAA911Ge22 Monoclonal Antibody to Aldosterone (ALD) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MEA911Ge Mini Samples ELISA Kit for Aldosterone (ALD) Enzyme-linked immunosorbent assay for Antigen Detection.
CEA911Ge ELISA Kit for Aldosterone (ALD) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA911Ge CLIA Kit for Aldosterone (ALD) Chemiluminescent immunoassay for Antigen Detection.
LMA911Ge Multiplex Assay Kit for Aldosterone (ALD) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
GUESTBOOK