High Sensitive ELISA Kit for Transferrin (TF)

TRF; Siderophilin; Serotransferrin; Beta-1 metal-binding globulin

  • High Sensitive ELISA Kit for Transferrin (TF) Packages (Simulation)
  • High Sensitive ELISA Kit for Transferrin (TF) Packages (Simulation)
  • High Sensitive ELISA Kit for Transferrin (TF) Results demonstration
  • HEC036Mu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the High Sensitive ELISA Kit for Transferrin (TF)

This assay has high sensitivity and excellent specificity for detection of High Sensitive Transferrin (TF).
No significant cross-reactivity or interference between High Sensitive Transferrin (TF) and analogues was observed.

Recovery of the High Sensitive ELISA Kit for Transferrin (TF)

Matrices listed below were spiked with certain level of recombinant High Sensitive Transferrin (TF) and the recovery rates were calculated by comparing the measured value to the expected amount of High Sensitive Transferrin (TF) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 86-101 97
EDTA plasma(n=5) 85-99 94
heparin plasma(n=5) 81-105 89

Precision of the High Sensitive ELISA Kit for Transferrin (TF)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Transferrin (TF) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Transferrin (TF) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the High Sensitive ELISA Kit for Transferrin (TF)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of High Sensitive Transferrin (TF) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 89-103% 90-97% 90-101% 81-89%
EDTA plasma(n=5) 94-103% 80-92% 93-101% 87-101%
heparin plasma(n=5) 97-104% 84-98% 86-102% 78-95%

Stability of the High Sensitive ELISA Kit for Transferrin (TF)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the High Sensitive ELISA Kit for Transferrin (TF)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the High Sensitive ELISA Kit for Transferrin (TF)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to High Sensitive Transferrin (TF). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to High Sensitive Transferrin (TF). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain High Sensitive Transferrin (TF), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of High Sensitive Transferrin (TF) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Mus musculus (Mouse) Applications (RESEARCH USE ONLY!)
NPC036Mu01 Native Transferrin (TF) Positive Control; Immunogen; SDS-PAGE; WB.
RPC036Mu01 Recombinant Transferrin (TF) Positive Control; Immunogen; SDS-PAGE; WB.
RPC036Mu02 Recombinant Transferrin (TF) Positive Control; Immunogen; SDS-PAGE; WB.
PAC036Mu01 Polyclonal Antibody to Transferrin (TF) WB; IHC; ICC; IP.
PAC036Mu02 Polyclonal Antibody to Transferrin (TF) WB; IHC; ICC; IP.
LAC036Mu81 FITC-Linked Polyclonal Antibody to Transferrin (TF) WB; IHC; ICC; IF.
LAC036Mu71 Biotin-Linked Polyclonal Antibody to Transferrin (TF) WB; IHC; ICC.
HEC036Mu High Sensitive ELISA Kit for Transferrin (TF) Enzyme-linked immunosorbent assay for Antigen Detection.
CEC036Mu ELISA Kit for Transferrin (TF) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC036Mu Multiplex Assay Kit for Transferrin (TF) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSC036Mu11 ELISA Kit DIY Materials for Transferrin (TF) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK