High Sensitive ELISA Kit for Interleukin 2 (IL2)

TCGF; Lymphokine; T-Cell Growth Factor; Aldesleukin

  • High Sensitive ELISA Kit for Interleukin 2 (IL2) Packages (Simulation)
  • High Sensitive ELISA Kit for Interleukin 2 (IL2) Packages (Simulation)
  • High Sensitive ELISA Kit for Interleukin 2 (IL2) Results demonstration
  • HEA073Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

This assay has high sensitivity and excellent specificity for detection of High Sensitive Interleukin 2 (IL2).
No significant cross-reactivity or interference between High Sensitive Interleukin 2 (IL2) and analogues was observed.

Recovery of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

Matrices listed below were spiked with certain level of recombinant High Sensitive Interleukin 2 (IL2) and the recovery rates were calculated by comparing the measured value to the expected amount of High Sensitive Interleukin 2 (IL2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-90 84
EDTA plasma(n=5) 81-96 91
heparin plasma(n=5) 91-104 96

Precision of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Interleukin 2 (IL2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Interleukin 2 (IL2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of High Sensitive Interleukin 2 (IL2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-101% 97-104% 87-98% 90-97%
EDTA plasma(n=5) 78-96% 83-95% 96-104% 89-104%
heparin plasma(n=5) 86-96% 87-101% 82-97% 96-103%

Stability of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the High Sensitive ELISA Kit for Interleukin 2 (IL2)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to High Sensitive Interleukin 2 (IL2). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to High Sensitive Interleukin 2 (IL2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain High Sensitive Interleukin 2 (IL2), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of High Sensitive Interleukin 2 (IL2) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
APA073Hu01 Active Interleukin 2 (IL2) Cell culture; Activity Assays.
APA073Hu02 Active Interleukin 2 (IL2) Cell culture; Activity Assays.
EPA073Hu61 Eukaryotic Interleukin 2 (IL2) Positive Control; Immunogen; SDS-PAGE; WB.
RPA073Hu01 Recombinant Interleukin 2 (IL2) Positive Control; Immunogen; SDS-PAGE; WB.
APA073Hu61 Active Interleukin 2 (IL2) Cell culture; Activity Assays.
RPA073Hu02 Recombinant Interleukin 2 (IL2) Positive Control; Immunogen; SDS-PAGE; WB.
PAA073Hu01 Polyclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
PAA073Hu06 Polyclonal Antibody to Interleukin 2 (IL2) WB
FAA073Hu01 Anti-Interleukin 2 (IL2) Polyclonal Antibody Flow cytometry.
LAA073Hu71 Biotin-Linked Polyclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC.
MAA073Hu26 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu21 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu22 Monoclonal Antibody to Interleukin 2 (IL2) WB
MAA073Hu24 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu25 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu23 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu27 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
MAA073Hu29 Monoclonal Antibody to Interleukin 2 (IL2) WB; IHC; ICC; IP.
HEA073Hu High Sensitive ELISA Kit for Interleukin 2 (IL2) Enzyme-linked immunosorbent assay for Antigen Detection.
MEA073Hu Mini Samples ELISA Kit for Interleukin 2 (IL2) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA073Hu ELISA Kit for Interleukin 2 (IL2) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA073Hu CLIA Kit for Interleukin 2 (IL2) Chemiluminescent immunoassay for Antigen Detection.
LMA073Hu Multiplex Assay Kit for Interleukin 2 (IL2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA073Hu01 ELISA Kit DIY Materials for Interleukin 2 (IL2) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK