CLIA Kit for Immunoglobulin G (IgG)

  • CLIA Kit for Immunoglobulin G (IgG) Packages (Simulation)
  • CLIA Kit for Immunoglobulin G (IgG) Packages (Simulation)
  • CLIA Kit for Immunoglobulin G (IgG) Results demonstration
  • CCA544Cp.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Immunoglobulin G (IgG)

This assay has high sensitivity and excellent specificity for detection of Immunoglobulin G (IgG).
No significant cross-reactivity or interference between Immunoglobulin G (IgG) and analogues was observed.

Recovery of the CLIA Kit for Immunoglobulin G (IgG)

Matrices listed below were spiked with certain level of recombinant Immunoglobulin G (IgG) and the recovery rates were calculated by comparing the measured value to the expected amount of Immunoglobulin G (IgG) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-101 98
EDTA plasma(n=5) 78-93 90
heparin plasma(n=5) 97-105 101

Precision of the CLIA Kit for Immunoglobulin G (IgG)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Immunoglobulin G (IgG) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Immunoglobulin G (IgG) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Immunoglobulin G (IgG)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Immunoglobulin G (IgG) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-94% 79-88% 82-101% 94-101%
EDTA plasma(n=5) 93-105% 79-103% 89-97% 78-94%
heparin plasma(n=5) 98-105% 87-94% 84-91% 98-105%

Stability of the CLIA Kit for Immunoglobulin G (IgG)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Immunoglobulin G (IgG)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
7. Read RLU value immediately.

Test principle of the CLIA Kit for Immunoglobulin G (IgG)

The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Immunoglobulin G (IgG). A competitive inhibition reaction is launched between biotin labeled Immunoglobulin G (IgG) and unlabeled Immunoglobulin G (IgG) (Standards or samples) with the pre-coated antibody specific to Immunoglobulin G (IgG). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Immunoglobulin G (IgG) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Immunoglobulin G (IgG) level in the sample or standard.

Related products

Catalog No. Organism species: Capra hircus; Caprine (Goat) Applications (RESEARCH USE ONLY!)
NPA544Cp01 Native Immunoglobulin G (IgG) Positive Control; Immunogen; SDS-PAGE; WB.
PAA544Cp01 Polyclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
SAA544Cp00 None-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Cp06 Cy3-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Cp09 HRP-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC.
SAA544Cp05 APC-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Cp04 PE-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Cp07 Biotin-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC.
SAA544Cp08 FITC-Linked Rabbit Anti-Caprine IgG Polyclonal Antibody WB; IHC; ICC; IF.
MAA544Cp25 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Cp27 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Cp21 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Cp22 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
SEA544Cp ELISA Kit for Immunoglobulin G (IgG) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA544Cp CLIA Kit for Immunoglobulin G (IgG) Chemiluminescent immunoassay for Antigen Detection.
LMA544Cp Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.