CLIA Kit for Galectin 1 (GAL1)

LGALS1; GBP; HLBP14; Galaptin; Lectin,Galactoside-Binding Soluble 1; 14 kDa laminin-binding protein; Lactose-binding lectin 1; Putative MAPK-activating protein PM12; S-Lac lectin

  • CLIA Kit for Galectin 1 (GAL1) Packages (Simulation)
  • CLIA Kit for Galectin 1 (GAL1) Packages (Simulation)
  • CLIA Kit for Galectin 1 (GAL1) Results demonstration
  • SCA321Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Galectin 1 (GAL1)

This assay has high sensitivity and excellent specificity for detection of Galectin 1 (GAL1).
No significant cross-reactivity or interference between Galectin 1 (GAL1) and analogues was observed.

Recovery of the CLIA Kit for Galectin 1 (GAL1)

Matrices listed below were spiked with certain level of recombinant Galectin 1 (GAL1) and the recovery rates were calculated by comparing the measured value to the expected amount of Galectin 1 (GAL1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-95 89
EDTA plasma(n=5) 78-104 91
heparin plasma(n=5) 79-101 97

Precision of the CLIA Kit for Galectin 1 (GAL1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Galectin 1 (GAL1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Galectin 1 (GAL1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Galectin 1 (GAL1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Galectin 1 (GAL1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 78-89% 84-92% 84-105%
EDTA plasma(n=5) 93-101% 90-98% 82-91% 87-102%
heparin plasma(n=5) 81-99% 80-90% 96-103% 97-105%

Stability of the CLIA Kit for Galectin 1 (GAL1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Galectin 1 (GAL1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Galectin 1 (GAL1)

The microplate provided in this kit has been pre-coated with an antibody specific to Galectin 1 (GAL1). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Galectin 1 (GAL1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Galectin 1 (GAL1) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
APA321Hu01 Active Galectin 1 (GAL1) Cell culture; Activity Assays.
RPA321Hu01 Recombinant Galectin 1 (GAL1) Positive Control; Immunogen; SDS-PAGE; WB.
PAA321Hu01 Polyclonal Antibody to Galectin 1 (GAL1) WB,IHC
FAA321Hu01 Anti-Galectin 1 (GAL1) Polyclonal Antibody Flow cytometry.
LAA321Hu71 Biotin-Linked Polyclonal Antibody to Galectin 1 (GAL1) WB; IHC; ICC.
LAA321Hu81 FITC-Linked Polyclonal Antibody to Galectin 1 (GAL1) WB; IHC; ICC; IF.
MAA321Hu22 Monoclonal Antibody to Galectin 1 (GAL1) WB; IHC; ICC; IP.
MAA321Hu21 Monoclonal Antibody to Galectin 1 (GAL1) WB
MEA321Hu Mini Samples ELISA Kit for Galectin 1 (GAL1) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA321Hu ELISA Kit for Galectin 1 (GAL1) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA321Hu CLIA Kit for Galectin 1 (GAL1) Chemiluminescent immunoassay for Antigen Detection.
LMA321Hu Multiplex Assay Kit for Galectin 1 (GAL1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA321Hu01 ELISA Kit DIY Materials for Galectin 1 (GAL1) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK