CLIA Kit for C Reactive Protein (CRP)

C-RP; PTX1; Pentraxin-Related; Pentraxin 1

  • CLIA Kit for C Reactive Protein (CRP) Packages (Simulation)
  • CLIA Kit for C Reactive Protein (CRP) Packages (Simulation)
  • CLIA Kit for C Reactive Protein (CRP) Results demonstration
  • SCA821Ra.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for C Reactive Protein (CRP)

This assay has high sensitivity and excellent specificity for detection of C Reactive Protein (CRP).
No significant cross-reactivity or interference between C Reactive Protein (CRP) and analogues was observed.

Recovery of the CLIA Kit for C Reactive Protein (CRP)

Matrices listed below were spiked with certain level of recombinant C Reactive Protein (CRP) and the recovery rates were calculated by comparing the measured value to the expected amount of C Reactive Protein (CRP) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-93 83
EDTA plasma(n=5) 80-97 91
heparin plasma(n=5) 95-104 98

Precision of the CLIA Kit for C Reactive Protein (CRP)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level C Reactive Protein (CRP) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level C Reactive Protein (CRP) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for C Reactive Protein (CRP)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of C Reactive Protein (CRP) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 92-102% 84-91% 87-101%
EDTA plasma(n=5) 98-105% 78-96% 97-105% 88-97%
heparin plasma(n=5) 87-95% 78-88% 82-103% 85-102%

Stability of the CLIA Kit for C Reactive Protein (CRP)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for C Reactive Protein (CRP)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for C Reactive Protein (CRP)

The microplate provided in this kit has been pre-coated with an antibody specific to C Reactive Protein (CRP). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to C Reactive Protein (CRP). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the C Reactive Protein (CRP) level in the sample or standard.;

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
RPA821Ra01 Recombinant C Reactive Protein (CRP) Positive Control; Immunogen; SDS-PAGE; WB.
APA821Ra01 Active C Reactive Protein (CRP) Cell culture; Activity Assays.
PAA821Ra01 Polyclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC; IP.
LAA821Ra71 Biotin-Linked Polyclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC.
LAA821Ra81 FITC-Linked Polyclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC; IF.
MAA821Ra21 Monoclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC; IP.
MAA821Ra22 Monoclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC; IP.
LAA821Ra82 FITC-Linked Monoclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC; IF.
LAA821Ra72 Biotin-Linked Monoclonal Antibody to C Reactive Protein (CRP) WB; IHC; ICC.
SEA821Ra ELISA Kit for C Reactive Protein (CRP) Enzyme-linked immunosorbent assay for Antigen Detection.
HEA821Ra High Sensitive ELISA Kit for C Reactive Protein (CRP) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA821Ra CLIA Kit for C Reactive Protein (CRP) Chemiluminescent immunoassay for Antigen Detection.
LMA821Ra Multiplex Assay Kit for C Reactive Protein (CRP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA821Ra01 ELISA Kit DIY Materials for C Reactive Protein (CRP) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK